Detail KYORITSU 1061/1062
KEW 1061
KEW 1062
Detection mode
RMS
MEAN/RMS (switch)
DC V
50.000/500.00/2400.0mV/5.0000/50.000/500.00/1000.0V
(Input impedance: Approx. 100MΩ [50/500/2400mV], 10MΩ [5/50/500/1000V])
|±0.02%rdg±2dgt (Basic accuracy)
AC V
[RMS]
50.000*/500.00mV/5.0000/50.000/500.00/1000.0V(* Only for 1062)
(Input impedance: 11MΩ<50pF [50/500mV/5V], 10MΩ<50pF [50/500/1000V])
|±0.7%rdg±30dgt (Basic accuracy)
|±0.4%rdg±30dgt (Basic accuracy)
AC V
[MEAN] *
--
50.000/500.00mV/ 5.0000/50.000/500.00/1000.0V
(Input impedance:
11MΩ<50pF [50/500mV/5V],
10MΩ<50pF [50/500/1000V])
|±1%rdg±30dgt (Basic accuracy)
DCV+ACV
5.0000/50.000/500.00/1000.0V
(Input impedance: 11MΩ<50pF [5V], 10MΩ<50pF [50/500/1000V])
|±1%rdg±10dgt (Basic accuracy)
|±0.5%rdg±10dgt (Basic accuracy)
DC A
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±0.2%rdg±5dgt (Basic accuracy)
AC A
[RMS]
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1%rdg±20dgt (Basic accuracy)
|±0.75%rdg±20dgt
|(Basic accuracy)
AC A
[MEAN] *
--
500.00/5000.0µA/50.000/500.00mA/ 5.0000/10.000A
|±1.5%rdg±20dgt (Basic accuracy)
DCA+ACA
500.00/5000.0µA/50.000/500.00mA/5.0000/10.000A
|±1.5%rdg±10dgt (Basic accuracy)
|±1%rdg±10dgt (Basic accuracy)
Ω
500.00Ω/5.0000/50.000/500.00kΩ/5.0000/50.000MΩ
|±0.1%rdg±2dgt (Basic accuracy)
|±0.05%rdg±2dgt (Basic accuracy)
LowPower-Ω *
--
5.000/50.00/500.0kΩ/5.000MΩ
|±0.2%rdg±3dgt (Basic accuracy)
Continuity buzzer
500.0Ω (The buzzer turns on for resistances lower than 100±50Ω)
Diode test
2.4000V|±1%rdg±2dgt
Open curcuit voltage: <5V (Approx. 0.5mA Measuring Current)
Capacitance
5.000/50.00/500.0nF/5.000/50.00/500.0µF/5.000/50.00mF
|±1%rdg±5dgt (Basic accuracy)
Frequency
2.000 - 9.999/9.00 - 99.99/90.0 - 999.9Hz/0.900 - 9.999/9.00 - 99.99kHz
|±0.02%rdg±1dgt (Basic accuracy)
DUTY
10 - 90%|±1%rdg
Temperature
-200 - 1372ºC|±1%rdg±1.5ºC
-328 - 2501.6ºF|±1%rdg±2.0ºF
(with the use of K-type Temperature probe)
Measurement
functions
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel (dBV, dBm), Continuity Check, Diode Test, LowPower-Ω *
Other functions
Data Hold (D/H), Auto Hold (A/H), Peak Hold * (P/H), Range Hold (R/H), Maximum value (MAX), Minimum value (MIN), Average value (AVG), Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory, LCD backlight
Withstand voltage
6,880V/5 sec.
Applicable
standards
IEC 61010-1 CAT IV 600V, CAT III 1000V Pollution degree 2,
IEC 61010-031, IEC 61326-1 (EMC)
Power source
R6 (1.5V)×4 (Auto power off: approx. 20 minutes)
Dimensions
192(L)×90(W)×49(D) mm
Weight
Approx. 560g (including batteries)
Accessories
7220A (Test leads), R6×4, Instruction manual,
8926 (Fuse [440mA/1000V])×1 (included)
8927 (Fuse [10A/1000V])×1 (included)
Tampilkan Lebih Banyak