Product Specification
Description
Dissolved Oxygen
Range: 0.00 to 90.00 mg/L or ppm
Resolution: 0.01 mg/L or ppm
Accuracy: ±0.2 mg/L
% Saturation
Range: 0.0 to 600.0%
Resolution: 0.1%
Accuracy: ±2 %
Temperature
Range: 0.0 to 50.0 °C (32 to 122 °F)
Resolution: 0.1 °C (0.1 °F)
Accuracy: ±0.3 °C (±0.5 °F)
Compensation: ATC / MTC (0.0 to 50 °C)
Salinity Correction
Range: 0.0 to 50.0 ppt
Resolution: 0.1 ppt
Method: Automatic correction after manual input
Pressure Correction
Range: 450 to 825mmHg (59.9 to 109.9 kPa)
Resolution: 1 mmHg (0.1 kPa)
Method: Automatic correction with inbuilt sensor
Calibration
2 Point (0%, 100%)
Operating Range
0.0 to 50.0 °C (32 to 122 °F)
Probe Type
Galvanic
Date/Time Stamping
Yes
Memory
500 data sets (GLP-compliant format)
Special Functions
Auto-Hold; Auto Power-Off after 2 - 30 minutes (user-preset); Stability Indicator; Self-diagnostics messages
Power
4 x 1.5 'AA' batteries (>400 hours) or 9V DC adapter, 500 mA